8 Start-up
FLUXUS F502TE 8.2 Switching on
67
UMFLUXUS_F502TEV1-5EN, 2020-05-29
• Select the temperature probe for the input T2.
• Press ENTER.
• Select heat for a heating application or chill for a cooling application.
• Press ENTER.
• Select return if the transducers are mounted on the return line or supply if they are mounted on the supply line.
• Press ENTER.
• Select sign if the sign of the thermal energy rate is to be considered or absolute if just the absolute value of the thermal
energy rate is to be displayed.
• Press ENTER.
• Select the input for reading the supply temperature. Select Fixed input val. to enter a fixed value for the supply
temperature or No measuring if no supply temperature is to be read.
• Press ENTER.
• Select the input for reading the return temperature. Select Fixed input val. to enter a fixed value for the return
temperature or No measuring if no return temperature is to be read.
• Press ENTER.
The current time is displayed.
• Press ENTER to confirm the time or set the current time via the keys and .
• Press ENTER.
The current date is displayed.
• Press ENTER to confirm the date or set the current date via the keys and .
• Press ENTER.
8.2 Switching on
As soon as the transmitter is connected to the power supply, the serial number of the transmitter is displayed for a short
time.
It is not possible to enter any data while the serial number is displayed.
Afterwards, the main menu is displayed in the default language. The language can be changed (see section 8.5).
Input T2
Thermal energy
Transd. Location
Thermal energy
T-Supply
T-Return
TIME
DATE