9 Measurement
9.3 Start of the measurement FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
84
• Select yes to repeat the test.
• Press ENTER.
9.2.5 Input of the error delay
The error delay is the time after which an error value will be sent to an output if no valid measured values are available.
This display will only be indicated if edit is selected in the menu item Special Funct.\Dialogs/Menus\
Error-val. delay.
If the error delay is not entered, the damping factor will be used.
• Enter a value for the error delay.
• Press ENTER.
9.3 Start of the measurement
• Select the program branch Measuring.
• Press ENTER.
If the parameters in the program branch Parameter are
not valid or incomplete, the error message NO DATA! will
be displayed.
If the transmitter has an RS485 interface or the data logger is activated, the measuring point number has to be entered:
Input of the measuring point number
• Enter the number of the measuring point.
• Press ENTER.
For the activation of text input, see Special Funct.\SYSTEM settings\Dialogs/Menus\Meas.Point No.
Output Options\...\I1:Test output?
Example
output: current output
output range: 4…20 mA
Zero-Scale Val.: 0 m³/h
Full-Scale Val.: 300 m³/h
Test value = 150 m³/h (center of the measuring range, corresponds to 12 mA)
If the external measuring instrument displays 12 mA, the current output functions correctly.
Output Options\...\I1:Error-val. delay
Measuring
Meas.Point No.
Sound Path
Transd. Distance
Signal test
Transd. Distance
Measurement display
see annex A, p. 149
Measuring\Channel\Meas.Point No.