Table 6. Error Messages
Message Description
batt
Appears on the primary display and indicates that
the battery is too low for reliable operation. The
Meter will not operate at all until the battery is
replaced. The B also appears when batt is on
the primary display.
bat
Appears on the secondary display and indicates
that the battery is too low to perform an insulation
test. The s button is disabled until the battery
is replaced. This message disappears when the
rotary switch is turned to any other function.
OPEn
Appears when an open thermocouple is detected.
LEAd
Test lead alert. The message appears briefly and a
single beep will sound when you move the switch
in or out of the c position.
15--
Err
Model detect error. Service Meter if this is
displayed.
di5c Meter cannot discharge a capacitor.
EPPr
Err
Invalid EEProm data. Have the Meter serviced.
CAL
Err
Invalid calibration data. Calibrate the Meter.
Input Terminals
Input terminals are shown and described in Table 7.
Table 7. Input Terminal Descriptions
.
1
2
3
4
bav04f.emf
Item Description
A
p
input terminal for insulation test.
B
q
input terminal for insulation test. Use for ac and dc
milliamp measurements to 400 mA and current
frequency measurements.
C
1577: input terminal for voltage, continuity, resistance
1587: input terminal for voltage, continuity, resistance,
diode, capacitance, voltage frequency, and
temperature measurements.
D
Common (return) terminal for all measurements except
insulation test.