Insulation Multimeter
General Specifications
31
General Specifications
Maximum Voltage Applied to
any Terminal and Common .................................... 1000 V
Fuse Protection for mA input .................................. Fast, 0.44 A, 1000 V, IR 10 kA
Batteries .................................................................. 4 AA IEC LR6 batteries, alkaline
Battery Life .............................................................. Meter use 1000 hours; Insulation test use: Meter can perform at least 1000 insulation tests with fresh
alkaline batteries at room temperature. These are standard tests of 1000 V into 1 MΩ with a duty cycle of 5
seconds on and 25 seconds off.
Size ......................................................................... 5.0 cm H x 10.0 cm W x 20.3 cm L (1.97 in H x 3.94 in W x 8.00 in L)
Weight ..................................................................... 550 g (1.2 lb)
Altitude
Operating ........................................................ 2000 m
Storage ........................................................... 12,000 m
Over-Range Capability ........................................... 110 % of range except for capacitance which is 100 %
Frequency Overload Protection .............................. 10
7
V Hz
Storage Temperature.............................................. -40 °C to 60 °C (-40 °F to 140 °F)
Operating Temperature .......................................... -20 °C to 55 °C (-4 °F to 131 °F)
Temperature Coefficient ......................................... 0.05 x (specified accuracy) per °C for temperatures <18 °C or >28 °C (<64 °F or >82 °F)
Relative Humidity .................................................... Noncondensing
0 % to 95 % @ 10 °C to 30 °C (50 °F to 86 °F)
0 % to 75 % @ 30 °C to 40 °C (86 °F to 104 °F)
0 % to 40 % @ 40 °C to 55 °C (104 °F to 131 °F)
Enclosure Protection............................................... IEC 60529: IP40 (non-operating)
Safety
IEC 61010-1 ................................................... Pollution Degree 2
IEC 61010-2-033 ............................................ CAT IV 600 V / CAT III 1000 V