Users Manual
36
Ohms Measurement
Range Resolution
Accuracy all 1587 models
[1]
+(% of Rdg+Counts)
Accuracy 1577
[1]
+(% of Rdg+Counts)
600.0 Ω 0.1 Ω
±(0.9 % + 2) ±(1.2 % + 2)
6.000 kΩ 0.001 kΩ
60.00 kΩ 0.01 kΩ
600.0 kΩ 0.1 KΩ
6.000 MΩ 0.001 MΩ
50.0 MΩ
[2]
0.01 MΩ ±(1.5 % + 3) ±(2.0 % + 3)
[1] Accuracies apply from 0 % to 100 % of range.
[2] Up to 80 % relative humidity.
Overload Protection ............................................ 1000 V rms or dc
Open Circuit Test Voltage ................................... <8.0 V dc
Short Circuit Current ........................................... <1.1 mA
Diode Test (all 1587 models)
Diode Test Indication .......................................... Display voltage drop: 0.6 V at 1.0 mA nominal test current:
Accuracy ............................................................. ±(2 % + 3)
Continuity Test
Continuity Indication ........................................... Continuous audible tone for test resistance below 25 Ω and off above 100 Ω. Maximum Reading; 1000 Ω
Open Circuit Voltage ........................................... <8.0 V
Short Circuit Current ........................................... 1.0 mA typical
Overload Protection ............................................ 1000 V rms
Response Time ................................................... >1 m sec
Frequency Measurement (all 1587 models)
Range Resolution
Accuracy ±(% of Rdg+Counts)
99.99 Hz 0.01 Hz
±(0.1 % + 1)
999.9 Hz 0.1 Hz
±(0.1 % + 1)
9.999 kHz 0.001 kHz
±(0.1 % + 1)
99.99kHz 0.01 kHz
±(0.1 % + 1)