374/375/376
Calibration Manual
12
1
4
7
/
+
STB
Y
5522A
CALIBRATOR
HI
LO
TRIG
GUARD
TC
20A
NORMAL
AUX
SCOPE
OUT
V, , ,
RTD
A, -SENSE,
AUX V
20V PK MAX
20V PK MAX
ghn50.eps
Figure 4. Current Calibration Setup
The calibration adjustment is complete.
When the adjustment is complete:
1. Remove the Power Supply.
2. Replace the batteries.
3. Reattach the battery compartment door.
4. Tighten the battery compartment door screw.