Front Panel Operation
Setting the Output
4
4-35
12. Press the numeric keys and decimal point key to enter the desired frequency output
(maximum five numeric keys). Press a multiplier key, if necessary. For example,
press the kilo multiplier key K. Then press the H key. For example,
1.1234 kHz.
13. The Control Display now shows your voltage and frequency entries. For example,
123.456 mV and 234.567 mV at 1.1234 kHz (below).
nn093f.eps
14. Press E. The calibrator clears your entry from the Control Display and copies
it into the Output Display (below is typical).
nn094f.eps
15. Press O to activate the calibrator output.
Two softkey labels appear on the Control Display: V@NOR/V@AUX and WAVE
MENUS.