EasyManua.ls Logo

Fluke 5520A - Leveled Sine Wave Function Verification: Frequency

Fluke 5520A
420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5520A-SC300 Option
Verification Tables
8
8-95
Leveled Sine Wave Function Verification: Flatness (cont.)
Nominal Value (p-p) Frequency Measured Value (p-p)
Deviation
(mV)
1-Year Spec.
(mV)
1.3 V 250 MHz 26.10
5.5 V 500 kHz 82.5
5.5 V 1 MHz 82.5
5.5 V 1 MHz 82.5
5.5 V 2 MHz 82.5
5.5 V 5 MHz 82.5
5.5 V 10 MHz 82.5
5.5 V 20 MHz 82.5
5.5 V 50 MHz 82.5
5.5 V 100 MHz 82.5
5.5 V 125 MHz 110.00
5.5 V 160 MHz 110.00
5.5 V 200 MHz 110.00
5.5 V 220 MHz 110.00
5.5 V 235 MHz 110.00
5.5 V 250 MHz 110.00
8-121. Leveled Sine Wave Function Verification: Frequency
Nominal Value (p-p) Frequency
Measured
Frequency Deviation 1-Year Spec.
1.3 V 50 kHz 0.0013 kHz
1.3 V 10 MHz 0.0003 MHz
1.3 V 250 MHz 0.0063 MHz