Remote Commands
Use of *OPC?, *OPC, and *WAI 6
6-17
Table 6-6. Command Summary (cont.)
5790B Calibration, Testing, and Diagnostics Commands
CAL_INTV? Returns the calibration interval.
CAL_NEXT Initiates the next calibration point.
CAL_NEXT?
Returns the value of the next point to be calibrated when in the middle of
calibrating a range.
CAL_OFF
Cancels any interactive calibration underway (that is, one started with the
CAL_AC, CAL_I2, or CAL_DC commands).
CAL_PASSWD Changes the calibration passcode.
CAL_RCSV?
Returns a comma-separated-value format report similar to what is saved on a USB
flash drive.
CAL_RPT? Returns a listing of a calibration report through the remote control port.
CAL_SECURE Locks/unlocks calibration security.
CAL_SECURE? Returns the lock state of calibration security.
CAL_SHIFT? Returns the calibration shift for a single calibration point.
CAL_USB Saves the calibration report to a USB flash drive.
CAL_SKIP
Advances to the next step of an interactive procedure without executing the step
(like a CAL_NEXT except the step is not executed).
CAL_SLST? Returns the calibration shifts for a group of service calibration points.
CAL_SPEC?
Returns the uncertainty specification for a specified input value, range, and
frequency.
CAL_STLST?
Lists the calibration constant groups that have been updated but not saved (and
therefore, how many would be written to nonvolatile memory by a CAL_STORE
command with the same argument).
CAL_STORE Stores calibration constants in the specified group list into nonvolatile memory.
CAL_STORE?
Lists how many calibration constant groups have been modified but not saved (this
is like CAL_STLST? except only the initial number is given).
CAL_TEMP
Sets the ambient temperature value stored when a calibration procedure is done.
Once set, the temperature is used for all calibration activities until changed.
CAL_TEMP? Returns the recorded ambient temperature of calibration.
CAL_WBCABLE Run the wideband cable calibration procedure.
CAL_WBLIN Start the interactive wideband linearity calibration procedure.
CAL_WBLIH? Lists the steps of the interactive wideband linearity calibration procedure.
CAL_ZERO Executes internal dc zeros calibration.
CMDSTR? Returns the command string that caused the last error.