Remote Commands
Alphabetical List of 5790B Device-Dependent Commands 6
6-25
AMPS?
Return corrected current shunt equivalent given the selected shunt.
Parameter
None
Response
1. (real) Amplitude
2. (String) Units
Example
AMPS?
1.1648312E+01,A
AMPS_DELTA?
Returns the shunt-corrected difference/ratio between the measurement and the
reference
Parameter
None
Response
1. (real) Amplitude
2. (String) PPM or PCT
Example
AMPS_DELTA?
+4.0E+00,PPM
AMPS_PKPK?
Returns the corrected current shunt peak-to-peak equivalent given the selected
shunt value and waveform type.
Parameter
SINE,SQUARE,TRI,TRUNC
Response
1. (real) Amplitude
2. (String) Units
Example
AMPS_PKPK? SINE
3.2830614E+00,A