Operators Manual
6-70
SUBNETMASK
Sequential command. Sets the Ethernet subnet mask for LAN communication
when NOT in DHCP mode.
Parameter
Subnet mask (quoted string consisting of 4 decimal values bound between 0-
255 separated by periods).
Example
SUBNETMASK "255.255.254.0"
Sets the Ethernet subnet mask to 255.255.254.0
SUBNETMASK?
Sequential command. Returns the Ethernet subnet mask for LAN
communication.
Parameter
None
Response
String
Example
SUBNETMASK?
Returns 255.255.254.0 if the subnet mask was previously set to this value.
TRIG
Triggers a measurement. When the Product is in manual trigger mode, a
measurement begins when you send the TRIG, *TRG, or MEAS? command. Use
the EXTRIG? query to determine the trigger mode setting.
Parameter
None
UNCERT?
Returns the uncertainty in ppm of the instrument for the present measurement
value and the present calibration interval setting. This returns 0 if there is no
specification for the present measurement.
Parameter
None
Response
1. (Float) The uncertainty in ppm
2. (CRD) PPM
3. (Integer) 90, 365, or 720, depending on the CAL_INTV setting
Example
UNCERT? 33.1,PPM,365