5790B
Operators Manual
B-2
Group FREQ
Name Default Function
FREQ_G 1.0 Frequency counter "gain" (crystal error)
Group DC_DAC
Name Default Function
DAC_G 3017.0 DAC gain
Group WDC_SENSOR
Name Default Function
SENSOR_C1_WB 3.162277660e-03 (Autogenerated)
SENSOR_C2_WB 0.0 (Autogenerated)
Group AC_LINEARITY
Name Default Function
LN_C .02 Coefficient of V^VEX*F^FEX
LN_LIM 100.0
Linearitization is done only below this
frequency
LN_VHI 2.0 Higher measured amplitude points
LN_VLO 0.6 Lower point
LN_CCALC 484.0 Use to figure LN_C (Vfs * F) ^ 2
Group RIPPLE
Name Default Function
RIP_LF 0.24 Multiplier with slow bit set
RIP_HF 25.6 Multiplier with slow bit not set