Figure 4-2. Recommended Test Lead Connections to Test a Source
External Guard Connection
For some conditions, such as low-frequency common mode interference, use the
external guard configuration for better results. Push
to select external guard.
The Guard indicator on the display is lit. This disconnects the LO from GUARD
and connects the internal guard to the GUARD binding post. A connection from
the GUARD binding post can then be made to the common earth ground point for
the system.