4-18
Current Shunt Function
Located in the Main Menu, the Product can be used with a current shunt to
determine the current-sourcing ability of a UUT by measuring the voltage output
of a current shunt as shown below. Refer to Instructions for Current
Measurements for information and instructions on how to use this function.
+/- 22.0 ppm (1 year) 83 days since calibration
2/06/15 10:52am
1.0000
0.800013
kHz
1.00003
1 A Shunt # 999002000
Input 2
Set
Reference
Setup
Menu
Peak-to-Peak Statistics
Auto Range 2.2 V
Select
Current
Shunt
hvi034.eps
Clearing the shunt returns the Product to normal operation.
Peak-to-Peak Function
In addition to the standard rms measurement, peak-to-peak voltage
measurements can also be shown. Push the Peak-to-Peak function on the
measurement screen to access the peak-to-peak measurement feature. Select
Sine, Square, Triangle, or Truncated sine waves to view the calculation on the
screen.