6270A
Operators Manual
16
PCM Installation
Depending upon how the Product is ordered, the PCM may come installed in the
unit or be delivered in a separate box.
To install the PCM:
Caution
To prevent damage to the sensors inside the modules, do not
drop the modules.
1. Remove the PCM from its shipping box.
2. Remove any protective plastic covers from the PCM pressure ports.
3. Confirm the O-rings are properly installed on each pressure port and are not
damaged.
4. Open the Module Bay as described in the Access the Module Bay section.
5. The PCM location in the Module Bay is on the far right. Align the track on the
bottom of the PCM with the PCM slot inside the Module Bay. See Figure 4.
6. Slide the PCM into place.
7. Tighten the two hex bolts on the front of the PCM. Torque to 0.5 N·m to
0.7 N·m (4 lbf · in to 6 lbf · in).
Caution
To prevent damage to the internal manifold, do not over torque.
8. Close and latch the front panel.
huo021.eps
Figure 4. PCM Installation