Table 7. Front-Panel Features (cont.)
Item Description
Use the Jog Wheel to make fine adjustments to the applied pressure. When turned,
the applied pressure changes by the least significant digit based on the
measurement resolution refer to Measurement Resolution for more information.
Jogging the pressure with the Jog Wheel adjusts the Setpoint value in any mode
but will only actively change the applied pressure while in Control mode. See the
subsequent section for more information on jogging pressure. Turn counter-
clockwise to decrease pressure or clockwise to increase.
Puts the Product into standby mode. In standby, the display is off and the keys are
disabled. Standby mode also disables remote operation. See Turn on the Product.
Emergency abort button that immediately vents the system pressure and stops all
pressure control. In addition, remote communication is stopped and the Product
enters a safety mode until manually deactivated. See Vent and Abort for more
information.
(ENTER)
Enter key used to accept and set numerical entries.