6270A
Operators Manual
34
Table 10. Control Settings
1
2
3
5
4
huo012.eps
Item Indicator/Name Function
Slew Rate
The slew rate is a user-configurable control parameter to adjust how
fast the Product controls pressure to the setpoint.
Control Limit
(Shown only when
Static is selected)
Related and used only with the Static Control mode, the control limit
is used to set an upper and lower pressure limit around the Setpoint.
The Product keeps the pressure within defined user limits. See Static
Control Mode and Control Limits (Only for Static Control).
Stability Limit
(Shown only when
Static is selected)
Related and used only for the Static Control mode, the stability limit
is a configurable tolerance to define the ready condition for the
measurement. The measurement indicator shows “Not Ready” until
the pressure is as stable as the value set in the stability limit. See
Stability Limits (Only for Static Control) for more information.
Ready Tolerance
(Shown only when
Dynamic is
selected. Not shown
in figure )
Related and used only with Dynamic Control mode, the Ready
Tolerance is a configurable band to define the ready condition for the
measurement. The band creates a positive and negative boundary
around the Setpoint pressure that is specified in pressure units or
%range of selected module. The measurement indicator shows “Not
Ready” until the pressure is stable within the band. See Ready
Tolerance for more information
Safety Limits
Protective user-configurable limits to protect the UUT from over-
pressure. An upper, lower, and vent limit can be manually set. See
Safety Limits.
Pressure Control
Mode
Shows and switches the active control mode between Dynamic and
Static. See Pressure Control Settings.