6270A
Operators Manual
74
Error Codes
If an error occurs during Product operation or control, an error message is shown
on the display. Errors can be caused by:
• Incorrect control that uses the front panel, (for example, attempts to force a
prohibited mode or overloading of output terminals)
• Product fault.
These messages are shown in Table 18. All error messages are shown in a
frame that overlaps the main screen.
Table 18. Error Codes
Error Number Error Message
0 No Error
103 Invalid Separator
104 Data Type
109 Missing Parameter
110 Command Header
113 Command Unknown
114 Header Suffix
201 Remote Required
221 Settings Conflict
222 Out of Range
281 Cannot create program
282 Illegal Program Name
284 Program Currently Running
285 Program Syntax Error
286 Program Runtime Error
313 Calibration Data Lost
315 Configuration Data Lost
330 Self-Test Failed
350 Queue Overflow
400 Query Error
500 Controller Malfunction
501 High Limit Exceeded
502 Low Limit Exceeded
503 Slew Limit Exceeded