Specifications
Specifications are based on a one year calibration cycle and
apply from +18 °C to +28 °C unless stated otherwise. Counts
means number of increments or decrements of the least
significant digit.
DC Volts Input
Range: + 28 V (+ 30 V max)
Input Impedance: 1 M
Accuracy: ± (0.025 % of reading + 2 counts)
DC mA Input
Range: 24 mA
Resolution: 0.001 mA
Accuracy: ± (0.02 % of reading + 2 counts)
DC mA Output
Range: 0 mA to 24 mA
Accuracy: ± (0.02 % of reading + 2 counts)
Source mode:
Compliance: 1000 at 20 mA
Simulate mode:
External loop voltage requirement: 24 V nominal, 30 V maximum,
12 V minimum
Loop Power
24 V
Percent display
-25 % to 125 %
Input/Output Protection
Input/output protection on mA range: resettable, nonreplaceable
0.1 A fuse.