719Pro
Calibration Manual
6
RTD Measure Verification
1. Push repeatedly until RTD is shown in the third row of the Product display.
2. Connect the 720URTDA to the Product.
3. Connect the 720URTDA Current HI to the 5522A HI RTD Output.
4. Connect the 720URTDA Sense HI to the 5522A HI RTD Output.
5. Connect the 720URTDA Current LO to the 5522A LO RTD Output.
6. Connect the 720URTDA Sense LO to 5522A LO RTD Output.
7. Set the 5522A to output the resistance setting in Table 5, and verify that the
display readings on the Product are within the limits.
Table 5. RTD Measure Verification Tests
Applied Current from 5522A Lower limit Upper Limit
84.270 Ω -40.10 °C -39.90 °C
100.000 Ω -0.10 °C 0.10 °C
157.325 Ω 149.90 °C 150.10 °C
8. Set 5522A to standby.
720RTD Probe Verification
Note
This is an optional test if the Product is paired with a 720RTD.
1. Push repeatedly until RTD is shown in the third row of the Product display.
2. Plug the 720RTD probe into the Product.
3. Insert the probe into the Metrology Well.
4. Set the 9170 to the temperatures in Table 6.
5. After the well stabilizes, verify that the display readings are within tolerance.
Table 6. 720RTD Probe Verification Tests
Applied Temperature Lower Limit Upper Limit
-30.00 °C -30.25 °C -29.75 °C
0.00 °C -0.25 °C 0.25 °C
70.00 °C 69.75 °C 70.25 °C
130.00 °C 129.75 °C 130.25 °C
6. Set 9170 to standby.