719Pro
Calibration Manual
4
4. Carefully vent all pressure and disconnect the first pressure port.
Table 2. Pressure Verification Tests
Input Pressure
6 Month Lower
Limit
6 Month Upper
Limit
12 Month
Lower Limit
12 Month
Upper Limit
719PRO-30G
30.000 29.992 30.008 29.989 30.011
25.000 24.992 25.008 24.989 25.011
17.500 17.492 17.508 17.489 17.511
12.000 11.992 12.008 11.989 12.011
5.000 4.992 5.008 4.989 5.011
0.000 -0.008 0.008 -0.011 0.011
-3.000 -3.008 -2.992 -3.011 -2.989
-7.000 -7.008 -6.992 -7.011 -6.989
-12.000 -12.008 -11.992 -12.011 -11.989
719PRO-150G
150.00 149.96 150.04 149.95 150.05
125.00 124.96 125.04 124.95 125.05
100.00 99.96 100.04 99.95 100.05
50.00 49.96 50.04 49.95 50.05
25.00 24.96 25.04 24.95 25.05
0.00 -0.04 0.04 -0.05 0.05
-3.00 -3.04 -2.96 -3.05 -2.95
-7.00 -7.04 -6.96 -7.05 -6.95
-12.00 -12.04 -11.96 -12.05 -11.95
719PRO-300G
300.00 299.92 300.08 299.89 300.11
250.00 249.92 250.08 249.89 250.11
175.00 174.92 175.08 174.89 175.11
100.00 99.92 100.08 99.89 100.11
50.00 49.92 50.08 49.89 50.11
0.00 -0.08 0.08 -0.11 0.11
-3.00 -3.08 -2.92 -3.11 -2.89
-7.00 -7.08 -6.92 -7.11 -6.89
-12.00 -12.08 -11.92 -12.11 -11.89