76
Service
1-6
Table 1-1. Specifications (cont)
Function Range Resolution Accuracy
Burden Voltage
(Typical)
400.0 mV 0.1 mV ±(1.9%+4)†
(45 Hz to 1 kHz) 4.000V 0.001V ±(1.9%+2)†
40.00V 0.01V ±(1.5%+2)† Not Applicable
400.0V 0.1V ±(1.5%+2)†
600V 1V ±(1.5%+2)†
4.000V 0.001V ±(0.3%+1)
40.00V 0.01V ±(0.3%+1) Not Applicable
400.0V 0.1V ±(0.3%+1)
600V 1V ±(0.3%+1)
40.00 mV* 0.01 mV ±(0.3%+5) Not Applicable
400.0 mV 0.1 mV ±(0.3%+1)
Ω 400.0Ω 0.1Ω ±(0.4%+2)
4.000 kΩ 0.001 kΩ ±(0.4%+1)
40.00 kΩ 0.01 kΩ ±(0.4%+1) Not Applicable
400.0 kΩ 0.1 kΩ ±(0.4%+1)
4.000 MΩ 0.001 MΩ ±(0.4%+1)
40.00 MΩ 0.01 MΩ ±(1%+3)
Capacitance 99.99 nF 0.01 nF ±(1.9%+2)**
999.9 nF 0.1 nF ±(1.9%+2)**
9.999 μF 0.001 μF ±(1.9%+2)** Not Applicable
99.99 μF 0.01 μF ±(1.9%+2)**
999.9 μF 0.1 μF ±(1.9%+2)**
9999 μF 1 μF ±10% Typical
400Ω 0.1Ω 5% Typical Not Applicable
40Ω 40Ω* 0.01Ω 5% Typical
(Lo-Ohms) 400Ω 0.1Ω 5% Typical Not Applicable
8 kΩ 1Ω 10% Typical
Diode Test 2.450V 0.001V ±2% Typical Not Applicable
4.000 mA 0.001 mA ±(1.5%+4)† 11 mV/mA
(45 Hz to 1 kHz) 40.00 mA 0.01 mA ±(1.5%+2)† 11 mV/mA
4A 0.001A ±(1.5%+4)† 0.03 V/A
10.00A*** 0.01A ±(1.5%+2)† 0.03 V/A
4.000 mA 0.001 mA ±(0.5%+5) 11 mV/mA
40.00 mA 0.01 mA ±(0.5%+2) 11 mV/mA
4A 0.001A ±(0.5%+5) 0.03 V/A
10.00A*** 0.01A ±(0.5%+2) 0.03 V/A
Frequency‡ 99.99 0.01 Hz ±(0.01%+1)
(1 Hz to 20 kHz) 999.9 0.1 Hz ±(0.01%+1) Not Applicable
9.999 kHz 0.001 kHz ±(0.01%+1)
20.00 kHz 0.01 kHz ±(0.01%+1)
* In 40Ω and 40 mV ranges, thermals may introduce additional errors. Maximum accuracy is obtained
when both probe tips are maintained at the same temperature.
** With film capacitor or better and residual (open lead reading) subtracted from measurement. This meter
uses a dc-type measurement technique.
*** 10A continuous, 20A for 30 seconds.
† From 5% to 100% of specified range.
‡ For rectangular waveforms 25% ≤ duty cycle ≤ 75% . VAC ≤ 1 kHz.