78
Service Manual
3-4
Table 3-1. Recommended Test Equipment
Equipment Minimum Specifications Recommended Model
DMM Calibrator DC Voltage: 0-500 V
Accuracy: .05%
AC Voltage: 0-500 V
Accuracy: 0.5%
Frequency: 100 Hz-20 kHz
DCA: 3.5-10 A
Accuracy: 0.2%
ACA: 10 A
Accuracy: 0.5%
Frequency: 45 Hz-1 kHz
Range: 0-35 MJ
Accuracy: 0.1%
Fluke Model 5500A
High Accuracy
Function Generator
Waveforms: Sine, Square &
Triangle
Voltage: 8 V peak
DC offset voltage: 0-4 V
Frequency: 10 Hz-20 kHz
Frequency Accuracy: .002%
Philips Models PM 5191, PM
5192, or PM 5193
DMM 100 mV DC Voltage Range
Accuracy: 0.5%
Resolution: 100 µ V
Fluke 20 Series, 70 Series or 80
Series
Thermometer Accuracy: 0.2 °C @ Ambient Mercury or Digital
Temperature Probe
Resolution: 0.1 °C or °F.
Must fit inside of a banana jack
Fluke 80T-150U
Lag Bath Room Temperature Water
For Testing the RPM-80 Inductive Pickup
Function Generator Single Cycle Triangle Waveform, -pi/2
Start Phase, 10 ms Repetition Rate
Frequency: 600 Hz
Output Voltage: 20 V (unterminated)
Philips Models PM 5133
or PM 5134
Oscilloscope 50 MHz, 3% vertical error 2 10X scope probes Philips PM 3055,
Fluke 93, 95, or 97
Adapter Binding Post to BNC Male Pomona Model 1296
Resistor 10 Ohm, ± 1.0% NA
Wire 14 or 16 AWG, 1 foot insulated NA