Appendix
Error Codes and Message A
AC Volts Tests
2801 100mV AC Range: Zero Input Check.
2811 100mV AC Range: +179mV DC Input Checks at Pre-Amp Output
2821 100mV AC Range: -179mV DC Input Checks at Pre-Amp Output
2831 1V AC Range: Zero Input Check.
2841 1V AC Range: +993mV DC Input Checks at Pre-Amp Output
2851 1V AC Range: -993mV DC Input Checks at Pre-Amp Output
2861 1V AC Range: +993mV DC Input Checks at RMS Converter Output
2871 1V AC Range: -993mV DC Input Checks at RMS Converter Output
2881 10V AC Range: Zero Input Check.
2891 10V AC Range: +9.78V DC Input Checks at Pre-Amp Output
2901 10V AC Range: -9.78V DC Input Checks at Pre-Amp Output
2911 100V AC Range: Zero Input Check.
2921 100V AC Range: +9.78V DC Input Checks at Pre-Amp Output
2931 100V AC Range: -9.78V DC Input Checks at Pre-Amp Output
2941 1000V AC Range: Zero Input Check.
2951 1000V AC Range: +9.78V DC Input Checks at Pre-Amp Output
2961 1000V AC Range: -9.78V DC Input Checks at Pre-Amp Output
A-5