EasyManua.ls Logo

Fluke 8845A - Page 76

Fluke 8845A
102 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8845A/8846A
Programmers Manual
68
Table 20. IEEE Common Commands (cont.)
Command Name Description
*IDN? Identification Query Meter returns the identification code of the meter as four fields
separated by commas. These fields are: manufacturer
("FLUKE"); model (“45"); seven-digit serial number; version of
main software and version of display software. (example:
FLUKE, 45, 9080025, 2.0, D2.0)
*OPC Operation Complete
Command
Meter sets the Operation Complete bit in the Standard Event
Status Register when parsed.
*OPC? Operation Complete
Query
Meter places an ASCII "1 " in the output queue when parsed.
*RST Reset Meter performs power-up reset, except that the state of IEEE
488 interface is unchanged, including: instrument address,
Status Byte, and Event Status Register.
*SRE <value>
Service Request
Enable
Sets the "Service Request Enable Register" to <value>, an
integer between 0 and 255. The value of bit six is ignored
because it is not used by the Service Request Enable
Register.
Note that <value> is an integer whose binary equivalent
corresponds to the state (1 or 0) of bits in the register. If
<value> is not between 0 and 255, an Execution Error is
generated.
*SRE?
Service Request
Enable Query
Meter returns the <value> of the Service Request Enable
Register (with bit six set to zero); <value> is an integer whose
binary equivalent corresponds to the state (1 or 0) of bits in the
register.
*STB? Read Status Byte Meter returns the <value> of the Status Byte with bit six as the
Master Summary bit.
<value> is an integer whose binary equivalent corresponds to
the state (1 or 0) of bits in the register.
*TRG Trigger Causes the meter to trigger a measurement when parsed.
*WAI Wait-to-continue Command required by IEEE 488.2 standard. Non-operational
in Fluke 45 Dual Display Multimeter. Command accepted but
has no effect.