8845A/8846A
Calibration Manual
A-8
Table A-1. Blank Verification Record for 90-Day Specifications (cont)
Nominal Output Test Limits
Ampl. Freq.
Range
High Low
Results Pass/Fail
1.0E-3 5000 1.0E-3 1.0014E-3 998.6E-6
1.0E-3 10000 1.0E-3 1.0045E-3 995.5E-6
8845A/8846A Current
10.0E-3 10 10.0E-3 10.014E-3 9.986E-3
10.0E-3 1000 10.0E-3 10.014E-3 9.986E-3
10.0E-3 5000 10.0E-3 10.014E-3 9.986E-3
10.0E-3 10000 10.0E-3 10.045E-3 9.955E-3
100.0E-3 10 100.0E-3 100.14E-3 99.86E-3
100.0E-3 1000 100.0E-3 100.14E-3 99.86E-3
100.0E-3 5000 100.0E-3 100.14E-3 99.86E-3
100.0E-3 10000 100.0E-3 100.45E-3 99.55E-3
1 45 1 1.0014E+0 998.6E-3
1 1000 1 1.0014E+0 998.6E-3
1 5000 1 1.0014E+0 998.6E-3
1 10000 1 1.0105E+0 989.5E-3
1.9 45 3 1.90465 1.89535
1.9 1000 3 1.90465 1.89630
1.9 10000 3 1.92765 1.87235
10 45 10 10.02100 9.97900
10 1000 10 10.02100 9.97900