EasyManua.ls Logo

Fluke 8846A - Using the Triggering Commands

Fluke 8846A
102 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Digital Multimeter
Supported SCPI Commands
59
SYSTem:DATE(?) <date>
Description Sets the system clock date.
Parameters <date> = MM/DD/YYYY or MM-DD-YYYY.
MM between 1 and 12
DD between 1 and 31
YYYY between 1970 and 2038
Examples SYST:DATE 10/25/2007
Query SYST:DATE? Returns the Meter’s system clock date.
SYSTem:TIME(?) <time>
Description Sets the system clock time in 24 format.
Parameters <time> = HH:MM:SS or HH-MM-SS
HH between 0 and 23
MM between 0 and 59
SS between 0 and 59
Examples SYST:TIME 14:25:10 Sets time to 2:25:10 PM
Query SYST:TIME? Returns the Meter’s system clock time.
SYSTem:VERSion?
Description Retrieves the Meter’s present SCPI command version.
Examples SYST:VERS? Returns SCPI version.
DATA:POINts?
Description Retrieves the number of measurements contained the Meter’s internal
memory.
Examples DATA:POIN? Returns number of measurements in internal
memory.
*RST
Description Resets the Meter to its power-up configuration.
Examples *RST Resets the Meter.
*IDN <n>, [<quoted string>]
Description Enables or disables the retrieval of a user-specified IDN string or factory-
set IDN string with the *IDN? command. Also sets the user defined IDN
string.
Parameters <n> = ON or 1 Enables user-defined string.
OFF or 0 Disables user defined string and causes the
factory IDN string to be returned for an *IDN?
command.
<quoted string> = Up to a 35 character string.
Example IDN ON, “My Meter” Sets the IDN string to My Meter and enables the
user defined string to be returned when the
*IDN? command is received by the Meter.