EasyManua.ls Logo

Fluke 8846A - Appendix C Network Consideration; Your Meter and Your Network Administrator

Fluke 8846A
112 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendices
Example Adjustment Program B
B-3
PRINT @<address of meter>, "CAL:VAL ZRES,10"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:REC"
## REAR CAL POINTS
PRINT @<address of meter>, "CAL:VAL ZRES,100000"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:VAL ZRES,10000"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:VAL ZRES,1000"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:VAL ZRES,100"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:VAL ZRES,10"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:VAL ZVDC,1"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:VAL ZVDC,0.1"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:VAL ZIDC,100.0E-3"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:VAL ZIDC,1.0E-3"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:VAL ZIDC,10.0E-3"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:VAL ZIDC,100.0E-6"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:REC"
## FRONT CAL POINTS
PRINT @<address of meter>, "CAL:VAL ZIAC,0.0"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:VAL ZIACS,0.0"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:VAL ZIAC,1.0E-3"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:VAL ZIACS,1.0E-3"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:VAL ZIAC,10.0E-3"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$

Table of Contents

Other manuals for Fluke 8846A

Related product manuals