8845A/8846A
Calibration Manual
B-4
PRINT @<address of meter>, "CAL:VAL ZIACS,10.0E-3"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:VAL ZIAC,100.0E-3"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:VAL ZIACS,100.0E-3"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:VAL ZIDC,10"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:VAL ZIDC,1"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:VAL ZIAC,1"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:VAL ZIACS,1"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:VAL ZIAC,10"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:VAL ZIACS,10"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:REC"
### Calibrate AC Linearity Set input value to 1.19V@1200Hz
# V@1200Hz
PRINT @<address of meter>, "CAL:VAL ACLIN,1.19"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
### Calibrate AC Linearity Set input value to 0.8V@1200Hz
# 0.8V@1200Hz
PRINT @<address of meter>, "CAL:VAL ACLIN,0.8"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
### Calibrate AC Linearity Set input value to 0.4V@1200Hz
# 0.4V@1200Hz
PRINT @<address of meter>, "CAL:VAL ACLIN,0.4"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
### Calibrate AC Linearity Set input value to 0.05V@1200Hz
# 0.05V@1200Hz
PRINT @<address of meter>, "CAL:VAL ACLIN,0.05"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
PRINT @<address of meter>, "CAL:REC"
### Calibrate 100 mV AC Gain @1200 Hz
# 100 mV AC Gain
PRINT @<address of meter>, "CAL:VAL GVAC,0.1"
PRINT @<address of meter>, “CAL? ON”
INPUT LINE @<address of meter>, A$
### Calibrate 100 mV AC Gain @1200 Hz