Introduction and Specifications
Electrical Specifications 1
1-19
8846A Accuracy
Accuracy is given as ± % measurement
Range Frequency
24 Hour
(23 ±1 °C)
90 Days
(23 ±5 °C)
1 Year
(23 ±5 °C)
Temperature
Coefficient/ °C
Outside 18 to 28 °C
3 – 5 Hz 0.1 0.1 0.1 0.005
5 – 10 Hz 0.05 0.05 0.05 0.005
10 – 40 Hz 0.03 0.03 0.03 0.001
40 Hz – 300 kHz 0.006 0.01 0.01 0.001
100 mV to
1000 V
[1][2]
300 kHz – 1 MHz 0.006 0.01 0.01 0.001
[1] Input >100 mV. For 10 – 100 mV, multiply percent measurement error by 10.
[2] Limited to 8 X 10
7
volt-Hertz
8845A Accuracy
Accuracy is given as ± % measurement
Range Frequency
24 Hour
(23 ±1 °C)
90 Days
(23 ±5 °C)
1 Year
(23 ±5 °C)
Temperature
Compensation/ °C
Outside 18 to 28 °C
3 – 5 Hz 0.1 0.1 0.1 0.005
5 – 10 Hz 0.05 0.05 0.05 0.005
10 – 40 Hz 0.03 0.03 0.03 0.001
100 mV to
750 V
[1][2]
40 Hz – 300 kHz 0.006 0.01 0.01 0.001
[1] Input >100 mV. For 10 – 100 mV, multiply percent measurement error by 10.
[2] Limited to 8 X 10
7
volt-Hertz
Gate Time vs. Resolution
Gate Time Resolution
0.01 5½
0.1 6½
1.0 6½