Users Manual
A-26
Table A-25. 96270A Microwave Output Sub and Inter-Harmonics Test
Output
Frequency
Sub/Inter-Harmonic Level
Frequency Actual Frequency Actual Frequency Actual Limit
4.05 GHz 6.075 GHz - - - - < -60dBc
6.2 GHz 9.300 GHz - - - - < -60dBc
8.3 GHz 4.150 GHz 6.225 GHz 12.450 GHz < -60dBc
12.0 GHz 3.000 GHz 6.000 GHz 9.000 GHz < -60dBc
16.0 GHz 4.000 GHz 8.000 GHz 12.000 GHz < -60dBc
18.5 GHz 4.625 GHz - - - - < -60dBc
22.00 GHz 5.500 GHz - - - - < -60dBc
27.00 GHz 6.750 GHz - - - - < -60dBc
Modulation - Microwave Output (96270A, Optional)
Use a high performance measuring receiver to make the measurements shown
in Tables A-26, A-27, A-28, and A-29.
Table A-26. 96270A Microwave Output Amplitude Modulation Rate Test
Level Frequency
Modulation
Rate
Depth Actual Tolerance
+10 dBm 30 MHz
20 kHz 50 % ±1.0 Hz
100 kHz 50 % ±10.0 Hz
Table A-27. 96270A Microwave Output Amplitude Modulation Depth Test
Level Frequency
Modulation
Rate
Depth Actual Tolerance
+14 dBm
125 MHz
1 kHz 90 % ±2.8 %
100 kHz 90 % ±2.8 %
1 GHz
1 kHz 90 % ±2.8 %
20 kHz 90 % ±2.8 %
Table A-28. 96270A Microwave Output Frequency Modulation Rate Test
Level Frequency
Modulation
Rate
Deviation Actual Tolerance
+13 dBm 62.5 MHz
400 Hz 30 kHz ±10 mHz
300 kHz 1 MHz ±10 Hz