DTX Series CableAnalyzer
Technical Reference Handbook
3-10
Figure 3-7. Autotest Summary Screen for Twisted Pair Cabling
amd77f.eps
A
PASS: All parameters are within limits.
FAIL: One or more parameters exceed the limit.
PASS*/FAIL*: One or more parameters are within the tester’s
accuracy uncertainty range, and the “*” notation is required by the
selected test standard. See “PASS*/FAIL* Results” on page
3-12.
B
Press K or L to scroll the screen.
C
If the test failed, press J for diagnostic information.
D
Action prompt for the screen. Use D A to highlight a
parameter; then press
H.
E
E: The measurement was within limits.
i: The parameter was measured, but has no PASS/FAIL limit in the
selected test limit. The results are for informational purposes only.
See pages 3-13, 3-17, and 3-18.
X: The measurement exceeds the limit.
U: See “PASS*/FAIL* Results” on page 3-12.
F
The worst margin found for the test.