DTX-MFM2/GFM2/SFM2 Fiber Modules
Users Manual
22
Autotest in Smart Remote Mode
1 Turn on the tester and smart remote and let them sit for
5 minutes. Allow additional time if the modules have
been stored above or below ambient temperature.
2 Turn the rotary switch to SETUP, then select Fiber. Set
the following on the Fiber tabs (press
C to see other
tabs):
Fiber Type: Select the fiber type to be tested.
Test Limit: Select the test limit required for the job.
Press
J More to see other lists of limits.
Remote End Setup: Set to Smart Remote.
Bi-directional: Enable this if you are required to test
the fiber in both directions.
Number of Adapters and Number of Splices: Enter
the number of adapters and splices that will be
added to each direction of the fiber path after the
reference is set.
Connector Type: Select the connector type used in
the cabling to be tested. Select General if the exact
type is not listed.
Test Method: Refers to the number of adapters
represented in the loss results. Select Method B if
you use the reference and test connections shown
in this manual.
3 Turn the rotary switch to SPECIAL FUNCTIONS, then
select Set Reference. If both a fiber module and a
twisted pair or coaxial adapter are attached, select Fiber
Module next.
4 The Set Reference screen shows reference connections
for the test method you selected. Figure 8 shows
connections for Method B. Clean the connectors on the
testers and reference test cords, connect the tester and
smart remote; then press
P.
See Table 2 on page 15 for details on settings.
-continued-