Documenting Process Calibrator HART Mode
Communication Operations
11
HART Output
From the Active Device screen press the Setup and HART
Output
softkeys to access the read/write HART Output
screen. Here you can change the Poll Address (0 = single
transmitter, any other address = multidrop), and control
burst mode.
pe12s.bmp
HART Information
From the Active Device screen press the Setup and HART
Information
softkeys to access the read-only HART
Information screen. This screen shows more complete
information about the transmitter model, hardware and
software revision numbers, and how many preambles it
sends.
pe13s.bmp