Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual
4-18
Input A, B, C, D Frequency Measurement Accuracy Test
Note
The test steps for channel C and D are only for the models
190-104, 190-204, and 190-504.
Proceed as follows to test the frequency measurement accuracy:
1. Connect the Test Tool to the 5502A as shown in Figure 4-5. Do not use 50 Ω
terminations.
NORMAL SCOPE
FLUKE 5502A CALIBRATOR
CONNECT TO CHANNEL A, B, C, D IN PARALLEL
50 OHM TERMINATIONNO
ADC
B
PM9093
PM9092
AB
METER
P
M
9
0
9
3
PM9092
PM9091
2 CHANN. + METER
4 CHANNELS
perf-ver-c2.eps
Figure 4-5. 5502A Scope Output to Test Tool Input A, B, C, D
2. Select the following Test Tool setup:
a) Recall the created setup (see Standard Test Tool Setup). Press
SAVE
,
F2
(RECALL) and select SETUP, press
ENTER
, select the setup name,
and press
ENTER
to recall the setup.
b) Press
SCOPE
.
c) Press
F2
(– READING …) and select with
F1
(– READINGS)
and with
:
Reading 1, on A, Hz
Reading 2, on B, Hz
Reading 3, on C, Hz
Reading 4, on D, Hz
3. Use
mV
RANGE
and
V
RANGE
to select range 100 mV/div for A, B, C and D.
4. Use
TIMEsns
to select the required time base setting.
5. Set the 5502A to source a sine wave according to the first test point in
Table 4-7.
Because no 50 Ω termination is applied, the 5502A leveled sine wave output
amplitude is twice the set value.
6. Observe reading
A, B, C, and D and check that it is within the range shown
under the appropriate column.