Multifunction PV Analyzer
Performance Tests
7
Figure 2. R
LO
Connections
8. Set the 532xA calibrator to output the first resistance test point (see Ta bl e 2 ) and set to
2-wire, ground bond resistance (Z GND) output mode.
9. Push
on the DUT to set the One Shot mode.
10. Push OPER on the 532xA to set the operate mode, if necessary, and wait for the calibrator
to stabilize.
Table 2 . R
LO
Te s t Po i n t s
Range
Te s t Po i n t s
[
Ω]
Display Reading [
Ω]
Lower Limit Upper Limit
0
Ω to 19.99 Ω
0.35 0.32 0.38
0.96 0.92 1.00
1.70 1.65 1.75
17.00 16.64 17.36
20
Ω to 199.0 Ω
90.0 88.0 92.0
170.0 166.4 173.6
200
Ω to 2000 Ω
470 445 496
900 853 947
1700 1613 1787
A
N
A
L
YZE
R
Fluke 5320A
Fluke SMFT-1000
LO
METER
,
L
,
L
+/-