R I N S E C Y C L E Continued
P/N 450368 Rev I 11
Inspect Spray Ball for correct operation:
Below Instructions is for Technical Service Use Only:
Figure 4: Spray Ball Ring Inspection/Cleaning Instructions
A. Remove the Small Access Port and look inside to check for
the Spray Ball Ring proper rotation.
B. Check the Spray Ball Ring for rotation.
SPRAY
BALL
RING
If Spray Ball Ring is clogged the water flow looks like
this, then call TECHNICAL SUPPORT (1-800-227-2572)
If Correct Spray and rotation of the Spray Ball Ring, then
proceed to step 4, pg 12.
D. Remove the nut below the Spray Ball to access the
Sprayer.
C. If the Spray Ball is stationary, push the PAUSE button.
RECIRCULATE Light will flash. Turn Main Power Switch OFF.
F. Turn main Power Switch ON. Press RINSE CYCLE START
button. PAUSE LED will turn off and RECIRCULATE LED will
not flash. Check for proper Spray Ring rotation before
continuing to pg. 12.
Ball a
E. Clear the spray holes in both the Spray Ball Ring and Sprayer
of debris and reassemble.