R I N S E C Y C L E Continued
12 P/N 450368 Rev I
4. When the RECIRCULATE Operation is complete, the GranuFlo
Dissolution Unit II will switch to DRAIN Operation and empty rinse
water from the GranuFlo Dissolution Unit II to the floor drain.
5. At the completion of the DRAIN Operation, the GranuFlo Dissolution
Unit II will refill to the 25 Gallon Sensor. The Fill Indicator Light will
turn ON and the second RINSE Operation will start.
6. When the second RINSE CYCLE is finished, the GranuFlo Dissolution
Unit II will go to the CYCLE COMPLETE Operation. The Rinse Cycle
Complete Indicator Light will turn ON and the Drain Valve will remain
OPEN, allowing any residual rinse water to go down the drain.