CAUTION! COMPARE THE CATALOG NUMBER ON THE STORAGE TANK WITH THE CATALOG NUMBER ON THE
GRANUFLO® DRY ACID CASE TO BE CERTAIN THEY ARE THE SAME.
1. After you obtained a valid specific gravity reading, remove the Transfer Hose from the
Transfer Nozzle.
2. Connect Transfer Hose to the storage tank and OPEN the Main Transfer Ball Valve at the side
of the Filter Housing. Press the Dissolution START button to transfer concentrate into the
storage tank (Figure 15).
Figure 15
3. Once Transfer of solution is complete, CLOSE the Main Transfer Valve, and place Transfer Hose
on to the GranuFlo Dissolution Unit II Transfer Hose Holder (see Figure 15). Do not leave
Transfer Hose attached to the Storage Tank.