7.3 TRANSFER TO INDIVIDUAL CONTAINERS
1. To transfer product into individual containers (containers must be properly labeled):
a. Connect the Transfer Hose to the
Transfer Nozzle.
b. Place the Transfer Nozzle into the
opening of an individual container.
c. Slightly OPEN the Main Transfer
Ball Valve on the top of the Filter
Housing.
d. With the Transfer Light flashing,
Press Dissolution START button.
The Transfer Pump will start.
e. Slowly OPEN Ball Valve on the
Transfer Nozzle until the desired
rate of flow through the nozzle is
achieved.
2. Once the manual TRANSFER Operation has been
completed and the individual containers are filled,
CLOSE Transfer Nozzle Valve. Press the PAUSE button.
3. When the GranuFlo Dissolution Unit II is empty, the
unit will step to the CYCLE COMPLETE Operation.
CLOSE the GranuFlo Dissolution Unit II Main Transfer
Ball Valve. Then Remove Transfer Nozzle and place
Transfer Hose onto the Transfer Hose Holder of the
Tank. Ensure you place the Transfer Nozzle in
appropriate area free from dirt and other
contaminates.
NOTE: IF THE PUMP IS RUN DRY FOR LONGER THAN 30 SECONDS IT MAY CAUSE PUMP FAILURE.
NOTE: ACID CONCENTRATE SHOULD NOT BE STORED IN MIX TANK LONGER THAN 14 DAYS FROM THE DATE OF
MIXING. ACID CONCENTRATE MAY BE STORED FOR LONGER PERIODS IN APPROVED CONTAINERS UNDER PROPER
STORAGE CONDITIONS.