2 P/N 450368 Rev I
SECTION 1: OVERVIEW
This product is manufactured by:
Fresenius USA, Inc.
4040 Nelson Avenue
Concord CA 94520
Fresenius USA, Inc. is a wholly owned subsidiary of Fresenius Medical Care Holdings, Inc.
The GranuFlo Mixer is automated and designed to mix a 375 liter (99 gallon) batch of dialysate concentrate with
a minimum of operator supervision, and is to be only used for the mixing of GranuFlo product.
There are two (2) pre-programmed cycles built into the Fresenius Medical Care GranuFlo Dissolution Unit II; the
RINSE CYCLE and the DISSOLUTION CYCLE. The RINSE CYCLE has four operations: FILL, RECIRCULATE, DRAIN,
and CYCLE COMPLETE. The first three: FILL, RECIRCULATE, and DRAIN are done twice before it reaches CYCLE
COMPLETE Operation. The DISSOLUTION CYCLE has eight operations: FILL, ADD GRANULES, MIX, DEAERATION,
FINAL FILL, HOMOGENIZE, TRANSFER, and CYCLE COMPLETE.
The control panel will display the GranuFlo Dissolution Unit II status at any given time, making it easy for the
operator to follow the progress of the operations in each cycle. The right side of the panel displays the RINSE
CYCLE, and the left side of the panel displays the DISSOLUTION CYCLE.
In addition, a manual PAUSE STATE has been incorporated into the unit. (See Section 9, page 27).
The following components are standard equipment for use with this device:
For other Original Equipment Parts see Spare Parts Manual, PN460009, on www.fmcna.com.