The hospital-grade plug for medical products intended for use in Denmark has DEMKO approval
and is rated 13 amps at 250Vac. Plug is recommended for use in medical applications and speci-
cations are being added to the standard SB 107-2-D1. Plug mates with maker’s Danish hospital-
grade socket. Hospital sockets have slightly di erent shaped openings allowing only the hospital
plug, not the standard Danish plug, to be inserted, to protect the ac circuit in speci c medical
settings.
A ground post, located on the back of the display, may be used for the purpose of grounding
the display’s chassis. Any such ground must be installed in accordance with applicable electrical
codes. The ground post is shown on the mechanical drawing found in this user’s guide.
Recycling
Follow local governing ordinances and recycling plans regarding the recycling or disposal of this
equipment.
Warning: Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment and the
other equipment should be observed to verify that they are operating normally.
Warning: Use of accessories, transducers and cables other than those speci ed or provided
by the manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
Warning: Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of this medi-
cal LCD monitor, including cables speci ed by the manufacturer. Otherwise, degradation of the
performance of this equipment could result.
Warning: Failure to use this equipment in the X-ray or magnetic resonance environment could
result in degradation of the performance of this equipment, interference with other equipment
or interference with radio services.
Warning: This equipment has been tested for radiated RF immunity only at selected frequencies,
and use nearby of emitters at other frequencies could result in improper operation.
Warning: This product is not considered physically to connect to HF (High Frequency) electrosur-
gical equipment.