108 INF-TN1FSC-E
10.4.2 Setting the output range
(1) Press the or key on the RANGE screen and
select “OUTPUT RANGE” from “OUTPUT”.
Press the
ENT
key, and the OUTPUT RANGE screen
is displayed
(1) Select “RANGE” on the OUTPUT RANGE screen
and press the
ENT
key and the cursor moves to
“KIND”.
(2) Set the kind of output range (velocity, fl ow rate and
thermal fl ow).
Press the
ENT
key, and the screen appears, enabling
you to select the output range.
Select the kind of output range by the
or key
and press the
ENT
key.
For the thermal fl ow of the range kind, make
10.3.3clause of "ENERGY MODE" "USED" and use
it.
100
%
Flow rate
ForwardReverse
Analog output
20mA
−100%
−20%
4mA
Output limit high
Output limit high
Output limit low
0 100%
200mA
4mA
Output limit low
Flow rate
Single range: Output one way by 0 to 100%
Output = Momentum flow rate ×100
/ Full scale
Bi-directional range: Also output back-flow by 0 to 100%.
When switching flow direction, Hysteresis
will be 10% in full scale.
(1) “RANGE”: when setting kind of output range, range type, full scale value
and output limit value.