INF-TN1FSC-E 149
State Cause Remedy
Measured value is not
zero though water ow
has stopped.
Water is subjected to convection in
a pipe.
Ö
Normal
Zero adjustment has been per-
formed.
Ö
Perform zero adjustment
again after making sure water
ow has stopped completely.
When water ow stops, pipe is not
lled up with water or becomes
empty.
Ö
The measured value, just
when ultrasonic wave cannot
be propagated, is held.
Ö
Normal
Measured value has an
error.
Input piping speci cations are dif-
ferent from actual ones.
Ö
A difference of 1% in inner
diameter causes an error of
about 3%.
Ø
• Input speci cations correctly.
Input scale as lining.
Because of an old pipe, scale has
stuck.
Ö
The length of straight pipe portion
is inadequate.
Ö
Select a different location of
sensor mounting (move the up-
stream side of a ow disturbing
element).
10D and 5D are required at least
on the upstream and downstream
sides.
Flow disturbing element should
not be present within 30D on
upstream side.
Pump, valve, ow joining pipe or
the like is unallowable.
Mount the sensor at differ-
ent angles with respect to the
cross section of pipe to ne
the location where mean value
is obtainable. The mount the
sensor at that location.
Pipe is not lled with water or mud
and sand have precipitated.
Ö
Precipitation is more when
the cross section of pipe has a
smaller area.
Ø
Shift the sensor to the vertical
portion of pipe.
Flow velocity pro le is
not available.
No re ector in the uid or weak
re ection.
Ö
Measure in a place subject to
re ector.
Low ow velocity.
Ö
Increase ow velocity.