INF-TN1FSC-E 23
(7) Display of mounting dimensions
After you nish the site setting on establish site screen, “Decision” is reversed from white to
blue by pressing
ENT
key.
Display the message “After sensor installation, please adjust Zero point”, turn back to “SITE
SETUP” screen.
At the last line the “SENSOR SPACING” value is displayed.
ENT
Mounting
dimension
Install the sensor according to chapter 8. MOUNTING OF DETECTOR and the mounting
dimension is as displayed on the last line.
For small pipe diameter, the sensor mounting length can be 0.0mm.
When the sensor mounting length is 0.0mm, error of the measurement is approxi-
mately ±2 to 5%.