6
INZ-TN2ZRE-E


Installation methods for the analyzer main unit are divided into 2 types;
Note) The analyzer weight should be supported at the bottom of the casing.
The analyzer should be installed in a place where ambient temperature is within -5 to 45°C (max.
40°C when two optical units are used, and the power supply is more than 200V AC), and tem-
perature uctuation during using is minimum.
Where vibration is unavoidable, protect the analyzer from vibrating.
For example, protection rubber is installed between the analyzer and support, or the analyzers
front panel and mounting panel.
“A” : 57.2 (EIA) or 50 (JIS)“A” : 57.2 (EIA) or 50 (JIS)
External dimensions Mounting dimensions Mounting method
(Unit : mm)
Type
19-inch rack
mounting
Panel cutout dimensions
Panel mounting
ESC
SPANMODE
ENTER
ZERO
GAS
ZRE
ANALYZER
ESC
SPANMODE
ENTER
ZERO
GAS
ZRE
ANALYZER
436
+2
-0
126
+2
-0
Mounting bracket
Support
Support
450 or more
465
A
M6
435
483
443
132.5
132.5
A