40
INZ-TN2ZRE-E
 
Auto calibration can be performed just once or forcibly stopped while the calibration is performed.

(1) In the “Setting of Auto Calibration”
screen that appears, point the cursor to
“Auto Calibration Run” by pressing the
or the key. Press the
ENT
key.
(2) “Run” is highlighted, displaying a mes-
sage to conrm the execution of auto
calibration. Press the
ENT
key to execute
the auto calibration, and press the
ESC
key to cancel.
This mode is used to stop the auto calibration forcedly.

(1) In the “Setting of Auto Calibration”
screen that appears, point the cursor to
“Auto Calibration Stop” by pressing the
or the key. Press the
ENT
key.
(“Auto Calibration Stop” appears when
the screen is selected while auto calibra-
tion is performed.)
(2) “Stop” is highlighted, displaying a mes-
sage to conrm the stop of auto calibra-
tion. Press the
ENT
key to stop the auto
calibration, and press the
ESC
key to
cancel (not stopped).