Fuji Electric ZRE-2 - Troubleshooting Specific Errors

Fuji Electric ZRE-2
85 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
69
INZ-TN2ZRE-E
In case of Error No. 5 and No. 7
• Pressing deletes the error display.
Calibration is continued. Unless another calibration
error occurs, calibration is carried out to the end, the
Measurement screen returns.
ENT
ENT
ESC
ESC
Calibrated
forcedly
C h
O2
0-25
vol%
009
.
C h
CO
0-200
ppm
00
C h
CO2
0-10
vol%
0000
C h
SO2
0-200
ppm
C h
NO2
0-25
ppm
908
.
.
.
136
.
Screen display and operation at the occurrence of error
In case of Error No. 1 to No. 4, No. 6, No. 8 to No. 10
ENT
ESC
ESC
ESC
• Press the key to delete the error display.
• If the key is pressed without removing
the cause of an error, the error will be
displayed again.
Measurement screen
When more than one error occurs, pressing
the key moves to another error display.
Display of error contents
C h
O2
0-25
vol%
2
100
.
C h
CO
0-200
ppm
00
C h
CO2
0-10
vol%
0000
C h
SO2
0-200
ppm
C h
NO2
0-25
ppm
008
.
.
.
.
Error No.9
136