• Protect the probes against severe vibration
• Do not kink the connector cable (risk of cable break)
• Never allow hard objects to contact rotating impellers.
• Always carry out probe cleaning according to the cleaning instructions -Never immerse probes in
solvent
• Never blow probes through with compressed air
• Allow hot high temperature probes to cool slowly; never cool by plunging them in cold water or
the like.
Instrument and probe must be switched off or disconnected prior to cleaning.
As the probes are highly sensitive measuring instruments, they must be cleaned with very great care.
Fibres or other foreign bodies can be carefully removed with fine tweezers.
When doing so, take care not to bend or otherwise damage the vanes or the spindle.
The adjustment of the bearing screws must never be changed.
This can result in erroneous measurements.