8
The Gas Data GFM has been designed to operate in typical field environments where flammable gases may be
present.
All ‘Special Conditions for Safe Use’ as detailed on the ATEX certificate (see Appendix F) must be adhered to.
The following points must be observed:-
The instrument must be recharged in the non-hazardous area.
The instrument must be checked for normal operation prior to carrying it into the hazardous
area:
o BEFORE entering a known hazardous zone, switch on the instrument and make sure
the display is visible and that the keys respond correctly.
o BEFORE entering a known hazardous zone, check that the sample pump can be
operated.
The instrument must be inspected for damage prior to use and the instrument must not be
used if the case is damaged:
o BEFORE entering a known hazardous zone, check the instrument for damage. Pay
particular attention to the keyboard and aspirator.
The instrument must be carried in the leather case to avoid the risk of electrostatic discharge:
o BEFORE entering a hazardous zone make sure that the instrument is securely enclosed
in its leather case.
_________________________________________________________________
Failure to comply the Special Conditions for Safe Usecan be potentially hazardous to the user and
others.
__________________________________________________________________
Warning